NOP-Nr:
Alternative:
5004
Overview
Instructions
Operating scheme
Substances
Equipment
Evaluation
Analytics
- Chromatogram
-
1
H-NMR
-
13
C-NMR
- IR
+
4-toluenesulfonic
acid
+
; Side reactions
Chromatogram
GC: crude product
column
Zebron ZB-1, L=15 m, d=0.25 mm, film 0.25 µm (Phenomenex)
inlet
injector: 250°C, split injection, 0.15 µL
carrier gas
He, pre-column pressure 101 kPa
oven
85°C (0 min), 8°C/min --> 250°C (10 min)
detector
FID, 250°C, H2 33.9 mL/min, synth. air 322 mL/min, isocratic N2, 15 mL/min (59 kPa)
integration
percent concentration calculated from relative peak area
GC: pure product
column
Zebron ZB-1, L=15 m, d=0.25 mm, film 0.25 µm (Phenomenex)
inlet
injector: 250°C, split injection, 0.15 µL
carrier gas
He, pre-column pressure 101 kPa
oven
85°C (0 min), 8°C/min --> 250° C (10 min)
detector
FID 250°C, H2 33.9 mL/min, synth. air 322 mL/min, isocratic N2, flow 15 mL/min (59 kPa)
integration
percent concentration calculated from relative peak area